Brief description
In combination with Rohde & Schwarz spectrum analyzers, the R&S®TS-EMF measurement system detects high-frequency electromagnetic fields in the environment (EMF). The isotropic antenna, together with the software, which has been specifically designed for EMF measurements, allows simple and precise evaluation of total and individual emissions on site.
Measuring EMF emissions in line with national and international standards provides a basis for ensuring that transmitter systems comply with the applicable limits and for communicating this to the general public. The key criterion is a precise, easy-to-make on-site measurement.
Key facts
features & benefits
Specifications
Isotropic antennas |
R&S®TSEMF-B3 |
R&S®TSEMF-B1 |
R&S®TSEMF-B2 |
Measurement principle |
isotropic reception due to orthogonally arranged antenna elements that are electronically switched |
||
Frequency range |
9 kHz to 200 MHz |
30 MHz to 3 GHz |
700 MHz to 6 GHz |
Minimum detectable field strength |
approx. 1 mV/m |
||
Maximum field strength |
≥100 V/m |
≥230 V/m |
≥230 V/m |
Isotropy |
≤ ±1.37 dB |
≤ ±2.1 dB, |
≤ ±2.5 dB (0.6 GHz to 2 GHz), |
Mechanical design |
radome protection against mechanical damage and environmental hazards |
||
Antenna factor |
individual calibration data, saved on USB dongle and/or CD |
||
Axis switching |
RF solid state switch |
||
Connecting cables |
direct connection to spectrum analyzer (extension cable (length: 8 m) see accessories) |
integrated cable (length: 2 m) ferritebeaded (extension cable (length: 8 m) see accessories) |
integrated cable (length: 2 m, extension cable (length: 8 m) see accessories)) |
RF connector |
N male |
||
Connector, control line |
7-pin connector (binder) for direct connection to R&S®FSH power sensor port, adapter cable for connection to notebook or spectrum analyzer with USB output |
||
Tripod adapter |
?-" thread, quick connector for antenna |
||
Requirements for notebook (not supplied) |
MS Windows7 or Windows8 operating system, free hard
disk space: min. 5 Gbyte |
||
Equipment supplied |
isotropic sensor with connecting cable operator’s guide for R&S®RFEX and R&S®RFEX-Fast measurement software (on CD), tripod adapter, USB adapter cable, transit case (R&S®TS-EMF with at least one isotropic antenna) |
||
Expanded measurement uncertainty R&S®TS-EMF with R&S®FSH (95 % confidence level (k = 2)) |
≤ ±2.5 dB |
≤ ±3.3 dB, |
≤ ±3.1 dB (0.7 GHz to 1 GHz), |
Expanded measurement uncertainty R&S®TS-EMF with R&S®FSL (95 % confidence level (k = 2)) |
≤ ±2.4 dB |
≤ ±3.2 dB, |
≤ ±2.9 dB (0.7 GHz to 1 GHz), |
Power supply |
via spectrum analyzer or laptop |
||
Ambient conditions |
–10 °C to +50 °C, safety class IP54 (see user manual for details) |
||
Weight (incl. cable) |
0.85 kg |
1.3 kg |
0.95 kg |
Dimensions (L × ?) |
550 mm × 146 mm |
475 mm × 170 mm |
415 mm × 87 mm |